Chapter 5 Preparation and Inspection of the System
5.1.2 Operation Check of Light Source
CAUTION
• To avoid damage to eyes, do not look at the light directly while the light is turned on. Do
not look directly at the light from the endoscope.
• While the light is turned on, do not look into the beam emitted from the distal end of
endoscope. Doing so may cause damage to the eyes.
Note
With regard the light seen through the ventilation holes, there are no safety problems.
(1) Turn on the power to the cart and the light source. The power indicator lamp lights up.
(2) Press the Pump button to make sure that the air supply pump’s operation switches in the
order of “HI,” “MID,” “LOW,” “OFF,” and “HI.”
(3) Turn ON the power to the processor.
The power indicator lamp lights up. The
. lamp of the EXAM. button lights up in
blue. The observation screen is displayed on the monitor.
FUJI TAROU
Procedure/Comment
Procedure/Comment
M45321
L-4+4
EG-L590ZW
10.8
9.8
BL7000_E-50_897N120551D.indb 4
2016/09/05 14:06:14
Summary of Contents for BL-7000
Page 2: ...ii 897N120551D BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 05...
Page 10: ...897N120551D 4 BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 06...
Page 18: ...897N120551D 1 8 Chapter 1 Precautions BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 06...
Page 52: ...897N120551D 6 10 Chapter 6 Method of Use BL7000_E 50_897N120551D indb 10 2016 09 05 14 06 17...
Page 66: ...897N120551D 8 6 Chapter 8 Troubleshooting BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 19...