897N120551D
8-1
Chapter 8
Trouble
shooting
Chapter
8
Troubleshooting
8.1 Troubleshooting
If the equipment does not work, or if any abnormality is found during operation, perform
inspections and take appropriate measures referring to the table below. If the phenomenon
persists, or it is not listed in the table below, stop using the equipment immediately and consult
your local FUJIFILM dealer.
Phenomenon
Possible cause
Countermeasure
Turning “ON” the power
button does not actuate
the equipment.
(The Power Indicator
Lamp does not light up.)
1) The main switch on the cart
is turned “OFF.”
2) The power cord is defective.
3) The receptacle is defective.
4) The fuse is blown.
1) Turn “ON” the main switch on the cart.
2) Check the power cord.
3) Check that the rated voltage is applied.
4) Consult your local FUJIFILM dealer.
The light does not turn
on even if the Light
button is pressed.
1) The endoscope is not
connected.
2) The LG detection sensor is
faulty.
3) The LED is faulty.
1) Securely lock the endoscope.
2) Consult your local FUJIFILM dealer.
3) Consult your local FUJIFILM dealer.
An image disappears
during examination.
1) The endoscope is not
connected completely.
2) The system is
malfunctioning due to
phenomena such as
electrostatic discharge.
3) The internal part of the light
source is faulty.
4) The system malfunctions
due to abnormal power
supply conditions (such as
voltage drop).
1) Withdraw the endoscope, ensure that
there is no foreign matter or dirt on
the power supply section, receiving
window or communication window, and
connect the endoscope again.
2) 3) 4)
Reset the processor and the light source.
If the image is not recovered, turn the
processor and the light source off,
straighten the bending section and then
pull out the endoscope slowly.
Stop using the equipment, and consult
your local FUJIFILM dealer.
Note
• Reset: Turn off the processor and the light source, and wait for at least 5 seconds. Turn
on the processor and the light source again, and then, turn on the light by pressing the
Light button.
• For details on how to operate the endoscope, refer to the operation manual of the
endoscope.
BL7000_E-50_897N120551D.indb 1
2016/09/05 14:06:19
Summary of Contents for BL-7000
Page 2: ...ii 897N120551D BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 05...
Page 10: ...897N120551D 4 BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 06...
Page 18: ...897N120551D 1 8 Chapter 1 Precautions BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 06...
Page 52: ...897N120551D 6 10 Chapter 6 Method of Use BL7000_E 50_897N120551D indb 10 2016 09 05 14 06 17...
Page 66: ...897N120551D 8 6 Chapter 8 Troubleshooting BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 19...