 Storage and Maintenance

7.3  Storage

7.3.1  Storage of Light Source

Note

  Do not store the equipment in a place which does not meet the storage conditions. 

Store the equipment in a state where no force is applied to the cables. The equipment 
may be damaged.

Store the BL-7000 system in a place that meets the following conditions.

<Storage Environment> 

Note

  

<Operating Environment> 

Non-operating conditions

Temperature:

 

-20 to +60°C

Humidity:

 

 

10 to 90%RH 

(no dew condensation)

Pressure:

 

70 to 106 kPa

Temperature:

 

-10 to +45°C

Humidity:

 

30 to 95%RH 

(no dew condensation)

Pressure:

 

70 to 106 kPa

When the equipment has been out of use for a long period, check the operation again as performed at 
the installation.

 

VP-7000 Operation Manual, “Chapter 6 Preparation and Inspection of the System”

7.3.2  Storage of Water Tank

For details on how to store the water tank, refer to the operation manual of the water tank.

BL7000_E-50_897N120551D.indb   3

2016/09/05   14:06:18

Summary of Contents for BL-7000

Page 1: ...ion of the Equipment and Peripherals 4 Preparation and Inspection of the System 5 Method of Use 6 Storage and Maintenance 7 Troubleshooting 8 Main Specification 9 BL 7000 Thank you for purchasing our...

Page 2: ...ii 897N120551D BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 05...

Page 3: ...Connection of the Equipment and Peripherals This chapter describes the installation and connection methods of the equipment and peripherals Chapter 5 Preparation and Inspection of the System This cha...

Page 4: ...4 1 2 4 Combination of Equipment 1 4 1 2 5 Electric Shock 1 4 1 2 6 Explosion 1 4 1 2 7 Maintenance 1 5 1 2 8 Preparation and inspection before use 1 5 1 2 9 Installation of Equipment 1 5 1 2 10 Hand...

Page 5: ...2 600 System Scope and 500 System Scope 5 2 5 1 2 Operation Check of Light Source 5 4 Chapter 6 Method of Use 6 1 6 1 Turning On the Light 6 2 6 2 Adjusting the Brightness 6 4 6 3 Operation of Air Su...

Page 6: ...Specification 9 1 Operating Environment 9 2 Transport and Storage Environment 9 2 Term of Validity Period for Use Durability 9 2 Input Output Connector 9 2 Block Diagram 9 2 Medical Device Directive...

Page 7: ...observe Light Source BL 7000 Operation Manual This manual provides necessary information for using the light source such as the equipment overview operation procedures and precautions to observe Note...

Page 8: ...able for malfunctions or damages resulting from negligence of the precautions and operating methods contained in this manual FUJIFILM Corporation shall not be liable for malfunctions or damages result...

Page 9: ...serious injury if not avoided CAUTION Explains situations that may cause slight or moderate levels of injury if not avoided Explains situations that may cause damage to equipment if not avoided 1 2 3...

Page 10: ...897N120551D 4 BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 06...

Page 11: ...res This manual does not provide information about clinical procedures or any aspects of endoscopic techniques CAUTION Do not modify this product or its components and do not disassemble repair or in...

Page 12: ...consult a cardiovascular specialist and the manufacturer of the active implantable medical device To avoid damage to eyes do not look at the light directly while the light is turned on Do not look dir...

Page 13: ...not rely on the BLI observation mode alone Perform comprehensive observation and diagnosis including normal observation for all target regions Not doing so may result in improper observation and diagn...

Page 14: ...of the distal end of the endoscope possibly resulting in mucosal burns or other injury Note Reset Turn off the processor and the light source and wait for at least 5 seconds Turn on the processor and...

Page 15: ...nspection result shows any abnormality do not use the equipment 1 2 9 Installation of Equipment CAUTION Since there are ventilation holes on the bottom surface of the light source BL 7000 take care no...

Page 16: ...s grease dust and gauze fibers adheres to the power receiving section If the connector mount is wet or soiled with foreign matter it may cause malfunction or failure of the devices Foreign matter wate...

Page 17: ...more of the following measures Change the orientation or position of any affected device Increase the spacing between devices Consult the manufacturer or dealer of the device Electromagnetic interfere...

Page 18: ...897N120551D 1 8 Chapter 1 Precautions BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 06...

Page 19: ...nts of BL 7000 The BL 7000 consists of the following items BL 7000 Components Note The operation manual this document is a component of BL 7000 The figure in parentheses indicates the quantity Light S...

Page 20: ...ired for general endoscopy Observation diagnosis and biopsy can be performed on the monitor Endoscope 700 System Scope 600 System Scope 500 System Scope Data Keyboard DK 7000U or DK 7000E Light Source...

Page 21: ...897N120551D 2 3 Chapter 2 Components and System Configuration of BL 7000 BL7000_E 50_897N120551D indb 3 2016 09 05 14 06 07...

Page 22: ...ension makes the following possible Endoscopic treatment Ultrasonic examination Recording of video images Printing of still images Data Keyboard DK 7000U or DK 7000E Processor VP 7000 Cart Light Sourc...

Page 23: ...P 7000 processor on the lower shelf and the BL 7000 light source on the upper shelf VP 7000 Operation Manual 4 1 6 Installation for System Expansion Electrosurgical Unit ICC 200 ERBE VIO300D ERBE LCD...

Page 24: ...897N120551D 2 6 Chapter 2 Components and System Configuration of BL 7000 BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 09...

Page 25: ...onnector or LG connector of the endoscope 2 Air Supply Indicator Displays the air supply pump s level of HI MID or LOW 3 Pump Button Switches the air supply amount in the four levels of HI MID LOW or...

Page 26: ...ghtness to be used as a standard 11 Light Mode Button Switches the Light mode to 1 2 3 or OFF 12 Light Mode Indicator Displays the Light mode 1 2 or 3 13 Light Limit Button Controls the light intensit...

Page 27: ...Terminal Connects to the equalization plug When necessary for safety this product is connected to a peripheral using this terminal to equalize the potential between the devices 3 Ventilation Holes Pr...

Page 28: ...897N120551D 3 4 Chapter 3 Name and Function of Each Part 3 3 Left Side Panel 1 1 Water Tank Hook Used to mount the water tank BL7000_E 50_897N120551D indb 4 2016 09 05 14 06 11...

Page 29: ...Function of Each Part 3 4 Bottom 2 1 3 1 Louver The filers are set 2 Filter Prevents intrusion of dust 3 Ventilation Holes Provide ventilation to keep the inside of the equipment cool BL7000_E 50_897N...

Page 30: ...897N120551D 3 6 Chapter 3 Name and Function of Each Part 3 5 Warning Labels Caution Label 3 6 Rating Plate Caution Label Rating Plate BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 12...

Page 31: ...t instructions for use Rear Panel Consult instructions for use Consult instructions for use Temperature limitation Keep dry CE marking Type BF applied part WEEE marking Humidity limitation Atmospheric...

Page 32: ...897N120551D 3 8 Chapter 3 Name and Function of Each Part BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 12...

Page 33: ...ventilation holes on the bottom surface of the light source BL 7000 take care not to cover the bottom surface with foreign substances Install equipment so that the power cord or connected endoscope ca...

Page 34: ...897N120551D 4 2 Chapter 4 Installation and Connection of the Equipment and Peripherals BL7000_E 50_897N120551D indb 2 2016 09 05 14 06 13...

Page 35: ...again 5 1 1 Installing the Endoscope and Water Tank CAUTION Before connecting the endoscope ensure that no foreign matter is caught in the scope connector section If any foreign matter is caught in t...

Page 36: ...tection cap CC 203 to protect the contacts The connecting method varies depending on the endoscope to be connected Turn OFF the processor or press the EXAM button to light up the STANDBY indicator lam...

Page 37: ...align it with the electrical connector index and then rotate the connector clockwise while lightly pressing it Note Make sure to rotate the connector index 90 degrees clockwise Do not connect more th...

Page 38: ...ith regard the light seen through the ventilation holes there are no safety problems 1 Turn on the power to the cart and the light source The power indicator lamp lights up 2 Press the Pump button to...

Page 39: ...center hole on the air water valve with a finger and make sure that air comes out of the air water nozzle Then take your finger off the hole and make sure that no air comes out of the air water nozzle...

Page 40: ...ve your hand toward and away from the distal end 8 Apply the palm of the hand to the distal end of endoscope and make sure that the image of a part of the hand can be seen Press the Light Mode Button...

Page 41: ...s operating normally If only one point of light or no light is seen stop using the equipment immediately turn off the equipment and consult your local FUJIFILM dealer Note If the light emitted from th...

Page 42: ...897N120551D 5 8 Chapter 5 Preparation and Inspection of the System BL7000_E 50_897N120551D indb 8 2016 09 05 14 06 15...

Page 43: ...e power button 6 1 Turning On the Light 4 Turn on the light 6 1 Turning On the Light 3 1 Front Panel 5 Adjust the brightness of the light 6 2 Adjusting the Brightness 6 Perform an endoscopic examinati...

Page 44: ...he light is turned on do not look into the beam emitted from the distal end of endoscope Doing so may cause damage to the eyes Note With regard the light seen through the ventilation holes there are n...

Page 45: ...e light turns on The pictogram for brightness level lights up in blue Note The light does not turns on when the scope connector or LG connector of the endoscope is not connected to the scope socket If...

Page 46: ...Brightness Adjustment button on the light source 1 When or is pressed the brightness level to be used as a standard can be adjusted Press to make the image brighter and press to make the image darker...

Page 47: ...Select HI MID or LOW 3 Press the air water valve of the endoscope to discharge water from the nozzle Note Be careful of the direction when water comes out 4 Close the center hole on the air water valv...

Page 48: ...aximum light intensity 1 Press the Transmitted Illumination button The lamp above the button flashes The light flashes with the maximum amount of the light 2 To reset the light intensity to the defaul...

Page 49: ...ome clotted in some cases even if the light save function is used 1 Press the Light Save button The maximum light intensity of the light is limited and the lamp above the Light Save button flashes 2 T...

Page 50: ...ress the Light Mode button The Light Mode changes in the order of 1 2 and OFF each time the Light Mode button is pressed 6 7 Turning Off the Light 1 Press the Light button at the left hand side of the...

Page 51: ...on on the front panel of the processor for 2 seconds or more Make sure that the STANDBY indicator lamp lights up in orange and then remove the endoscope Note Pull the scope connector straight out Note...

Page 52: ...897N120551D 6 10 Chapter 6 Method of Use BL7000_E 50_897N120551D indb 10 2016 09 05 14 06 17...

Page 53: ...sty or dirty wipe it gently with a soft cloth such as gauze To remove heavy dirt wipe the equipment gently with gauze moistened with a small quantity of neutral detergent solution diluted 10 times or...

Page 54: ...e 2 Remove dust from the filters of the louver Filters Louver 3 Attach the louver to the light source Make sure that the louver is fully inserted into the light source Not doing so may cause equipment...

Page 55: ...ating Environment Non operating conditions Temperature 20 to 60 C Humidity 10 to 90 RH no dew condensation Pressure 70 to 106 kPa Temperature 10 to 45 C Humidity 30 to 95 RH no dew condensation Pressu...

Page 56: ...oid hitting it against the scope connector or LG connector of the endoscope connected to the light source There is a risk of damaging the endoscope or light source When moving the light source install...

Page 57: ...pacts to the equipment paying close attention to bumps and slopes Note Use two persons to move the cart Pay close attention to prevent the monitor and the scope hanger from hitting anything 4 Lock the...

Page 58: ...ll out the endoscope slowly Note For details on how to operate the endoscope refer to the operation manual of the endoscope When the Light Source Status Lamp flashes when it is on If the alarm flashes...

Page 59: ...button to turn on the light If the light turns on the equipment is ready to operate If not follow the procedure in 3 3 Turn off the power of the light source and processor and turn on the power again...

Page 60: ...Light button If the light turns on the equipment is ready to operate 7 6 Cleaning and Disinfecting or Sterilizing the Water Tank For details on how to clean and disinfect or sterilize the water tank r...

Page 61: ...y lock the endoscope 2 Consult your local FUJIFILM dealer 3 Consult your local FUJIFILM dealer An image disappears during examination 1 The endoscope is not connected completely 2 The system is malfun...

Page 62: ...nd the light source off straighten the bending section and then pull out the endoscope slowly 3 Withdraw the endoscope ensure that there is no foreign matter or dirt on the receiving window or on the...

Page 63: ...apy device to restore image output The endoscope is working properly 2 Withdraw the endoscope ensure that there is no foreign matter or dirt on the receiving window or on the communication window and...

Page 64: ...supplied 1 The air supply pump is OFF 2 The water tank cover is not closed completely 3 Too much water in the water tank 4 The water tank is empty 5 It is not connected to the endoscope firmly 1 Press...

Page 65: ...on the BL 7000 2 Securely connect the interface cable Unplug and plug back endoscope reset the processor and the light source If the problem persists contact your local FUJIFILM dealer Cause The light...

Page 66: ...897N120551D 8 6 Chapter 8 Troubleshooting BL7000_E 50_897N120551D indb 6 2016 09 05 14 06 19...

Page 67: ...50 60Hz Current consumption rated 1 2 0 7A Fuse T4AH 250V 2 Rating 4A 250V Light control Automatic light control by the control signal from the processor Light cooling method Forced air cooling Air su...

Page 68: ...or 1 channel Block Diagram The main unit is provided with the power supply section control section operation switches on the front panel interface section LED drive section endoscope scope connector s...

Page 69: ...9 3 Chapter 9 Main Specification Medical Device Directive This product complies with the requirements of European Directive 93 42 EEC Classification Class II a BL7000_E 50_897N120551D indb 3 2016 09...

Page 70: ...nce RF emissions EN 55011 Group 1 This product uses RF Radio Frequency energy for its internal function Its RF emissions are very low and are not likely to cause any interference in nearby electric eq...

Page 71: ...environment Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 5 UT 95 dip in UT for 0 5 cycle Same as left Main power quality should be that of a typica...

Page 72: ...nsmitter Recommended separation distance d P V1 3 5 d P E1 3 5 80 to 800MHz d P E1 7 800MHz to 2 5GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmit...

Page 73: ...his product and cell phone or RF communication equipment table below Referring to the table below keep cell phone or RF communication equipment away from this product Rated maximum output power of tra...

Page 74: ...b Trouble caused by careless handling or misuse of the product on the part of the user c Trouble caused by repair or modification by an unauthorized person 4 Repairs after the warranty period We will...

Page 75: ...point for the recycling of electrical and electronic equipment By ensuring this product is disposed of correctly you will help prevent potential negative consequences to the environment and human hea...

Page 76: ...3 2 Light Save Indicator 3 2 Light Source Status Lamp 3 2 5 5 7 6 Louver 3 5 M Method of Use 6 2 O ON Indicator 3 1 P Potential Equalization Terminal 3 3 Power Button 3 2 power cord 3 3 Power Indicat...

Page 77: ...o locate our representative in your country USA Fujifilm Medical Systems U S A Inc http www fujifilmendoscopy com 800 385 4666 Australia FUJIFILM Australia Pty Ltd http www fujifilm com au 1800 060 20...

Page 78: ...Europe GmbH Heesenstrasse 31 40549 Duesseldorf Germany FUJIFILM Corporation 26 30 Nishiazabu 2 chome Minato ku Tokyo 106 8620 Japan EU Importer FUJIFILM Europe B V Oudenstaart 1 5047 TK Tilburg The Ne...

Reviews: