![Fuji Electric FE ZSS Instruction Manual Download Page 74](http://html1.mh-extra.com/html/fuji-electric/fe-zss/fe-zss_instruction-manual_2341881074.webp)
- 64 -
INZ-TN1ZSS-E
7. MAINTENANCE
7.1 Maintenance list
To maintain the desired accuracy, we recommend you to perform periodical maintenance and inspection,
referring to Table 7
Table 7
-
1
Maintenance cycle
Items
6 months
1 year
Light axis adjustment
○
Replacement of packing
○
Zero calibration
○
Span calibration
○
7.2 Maintenance procedure
To operate the instrument properly and keep it in favorable operation status, it is essential to perform main-
tenance and inspection periodically.
Note that the Table 7
2 provides the guideline for maintenance items and intervals, assuming standard gas,
operation, and installation environment. When you perform maintenance by yourself, add the items or mod-
ify intervals as required to suit the installation environment.
However, do not extend the maintenance cycle of each maintenance item.
We do not warrant the damage resulting from the use not observing periodical maintenance.
Table 7
-
2
Maintenance cycle
Locations Maintenance
items
6 months
1 year
Con-
tents
Control unit
Check that contents of error or error
alarm are not displayed.
c
↓
Receiver/
transmitter unit
Check that the equipment is securely
fixed to the stack and there is no vibra-
tion.
c
↓
Receiver/
transmitter unit
Zero calibration
c
P.34
↓
Receiver/
transmitter unit
Span calibration
c
P.37
↓
Receiver/
transmitter unit
Replacement of packing
c
P.65
↓
Receiver/
transmitter unit
Light axis adjustment
c
P.65
↓
Receiver/
transmitter unit
Check for air purge flow rate
c