INZ-TN1ZSS-E
- 45 -
6.4.1 Output hold
Set output hold to “Last Meas.” or “Pre-set” to hold analog output. (Indication value on the screen is not
held.)
Point the to the measurable component to
hold output by the
key or the
key,
and press the
key.
(
)
Select either “Last Meas.” or “Pre-set” by the
key or the
key. Press the
ENT
key
to validate the setting.
(
)
ENT
When “No” or “Last Meas.” is selected, cursor
returns to the starting position.
When “Pre-set” is selected, numeric value is
highlighted. Change the numeric value by the
key or the
key, and move the digit to
the right by the
key.
After the numeric value is changed, press the
ENT
key.
(
)
ENT
Output Hold
HC
ℓ
No
020
%FS
2007-4-1
12:00:00
Output Hold
HC
ℓ
Pre-set
020
%FS
2007-4-1
12:00:00
Output Hold
HC
ℓ
Pre-set
0
20
%FS
2007-4-1
12:00:00