Setting Range
“Last Meas.” : Holds the value for which “Last Meas.” is determined by the
ENT
key.
“Pre-set”
: Holds the %FS value for which range is currently validated. Previous set value is kept held
even if the range is switched during the hold.
Example) When range is from 0 to 10ppm, and the set value is 20%FS, the value corre-
sponding to 2ppm is output regardless of the measurement value.
O
2
Conversion Hold
“Last Meas.” : Holds the measurement value for which “Last Meas.” is determined by the
ENT
key, and
the value calculated by O
2
value. When the O
2
value is fixed, it holds the value calculated
by the fixed value. Even if O
2
analog input is entered, it holds the value calculated by O
2
value which is determined by the
ENT
key.
“Pre-set”
: Holds the value calculated by %FS value of the currently enabled range and O
2
value. The
O
2
value is the same as the last value.
Last Meas. of average value
When the average value output is selected and “Last Meas.” is clicked, the averaged value is kept holding
until determined by the
ENT
key.
Output Hold
HC
ℓ
Pre-set
020
%FS
2007-4-1
12:00:00