Sample Extractor Assemblies
FTS Systems LyoStar™ 3
32
Rev 006, 01/14
© SP Scientific 2014
Press down on the stoppering ram to manually stopper the vial.
9.
Before the sample can be removed from the antechamber, the antechamber
10.
pressure must be returned to atmospheric pressure. If you connected an inert
gas source for backfilling, remove it now. Open the vacuum break valve to
release vacuum from the antechamber. Allow the antechamber to return to
atmospheric pressure.
Loosen the thumbscrews on each side of the antechamber and push down on
11.
the front plate until it can be opened.
Note: Vacuum pressure in the chamber will cause the antechamber front plate gasket
to seal. In order to break the seal, you may need to apply force initially when sliding
down the front plate.
Flip down the front plate.
12.
Stoppering Ram
Stopper
Vial Sample
Thumbscrews
Antechamber
Front Plate
Push down on the front plate until it can be
opened. You may need to apply force initially.
Antechamber Front Plate
Summary of Contents for LYOSTAR 3
Page 2: ......
Page 6: ...iv Rev 006 01 14 SP Scientific 2014...
Page 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Page 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Page 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Page 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Page 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Page 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Page 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Page 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Page 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Page 81: ......