FTS Systems LyoStar™ 3
Getting Started
Rev 006, 01/14
5
© SP Scientific 2014
Clearance.
Twenty-four (24) inches (61 cm) of obtainable clearance is
7.
recommended on all sides of the unit for serviceability and efficient operation.
If multiple machines are being operated in a single area, position them so that
the hot exhaust from one does not blow into the intake of another. If machines
are placed side-by-side, increase the minimum clearance to 48 inches
(122 cm).
Cleanroom systems:
Position the freeze dryer so that the wall flange is flush against
the periphery of the wall opening. Screw, caulk and secure it to the wall.
Heat Transfer System.
The heat transfer system is shipped with heat
8.
transfer fluid. The fluid circuit is a sealed system and under normal setup or
operating conditions does not require additional fluid. The heat transfer fluid
level may be monitored through the sight glass located on the side of the heat
transfer fluid reservoir. This sight glass is visible through the upper rear grill
panel.
Note: The heat transfer fluid level should be approximately ¾ full when the system is
at 25 °C.
Door Gaskets.
Ensure that the chamber and condenser door gaskets are
9.
clean and free of defects (
e.g.
, cuts, cracks, tears). If the gasket shows any
signs of cracks or tears replace it immediately. Failure to do so shall result in
poor vacuum integrity. As a minimum, gaskets should be inspected quarterly
and replaced annually.
Note: The door gaskets do not require vacuum grease for pre-seal or operation.
Vacuum System
10.
Vacuum Pump Oil.
Check the vacuum pump oil. The vacuum pump is
shipped with oil in it. The vacuum pump oil should be clearly visible in the
sight glass on the front of the pump. The oil level should read between the
center and top of the sight glass while the pump is off and to the top of the
sight glass while the pump is in use. Add oil only if necessary. DO NOT
OVERFILL!
Note: For information on vacuum system maintenance, refer to Chapter 10: General
Maintenance, Vacuum System .
Oil Mist Eliminator.
The LyoStar™ 3’s vacuum system is equipped with an
Oil Mist Eliminator. Ensure that it is properly attached to the vacuum system
before operating the lyophilizer.
Oil Mist Eliminator (OME)
Vacuum Pump Oil Sight Glass
Summary of Contents for LYOSTAR 3
Page 2: ......
Page 6: ...iv Rev 006 01 14 SP Scientific 2014...
Page 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Page 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Page 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Page 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Page 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Page 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Page 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Page 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Page 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Page 81: ......