FTS Systems LyoStar™ 3
Sample Extractor Assemblies
Rev 006, 01/14
31
© SP Scientific 2014
Slide the isolation door closed and lower the handle to lock it into position.
7.
Remove the plastic shroud from the stoppering ram.
Prior to stoppering the sample, you may backfill the antechamber with an inert
8.
gas, such as nitrogen (N
2(gas)
) via the vacuum break valve. Connect your inert
gas source to the vacuum break valve via the filter attachment.
Plastic Shroud removed from
stoppering ram
Vacuum Break Valve
Isolation Door
Handle lowered to lock
isolation door into position.
Vacuum Break Valve
Connect an inert gas source
to the vacuum break valve via
the filter attachment.