FC Series w/ EDC Control
Thermal
MNL-047-A
Division Rev.
004
FTS
32
Stone Ridge, NY USA
845.687.0071
FSPANL? {F022}
Query the low temperature limit (degrees) established by the FLUID definition.
GNREM= / GNREM? {F023}
Adjust or query the calibration gain factor applied to the remote [external] temperature sensor.
GNRTD= / GNRTD? {F024}
Adjust or query the calibration gain factor applied to the system [internal] temperature sensor.
HEATER? (Service Use) {F025}
Query the system heater configuration and phase connection definition.
HLPC? (Service Use) {F026}
Query the existence of high-low-pressure cutout on the system compressor(s). Used in water-cooled units.
0 = not used, 1 = exists.
HPB= / HPB? {F027}
Set or query the width, in degrees, of the band about setpoint (SP) over which the system will proportion
the power applied to the heater (Heat Proportioning Band).
HYSTLI? (Service Use) {F028}
Query the hysteresis, in degrees, applied to activation of the liquid injection valve.
HYSTST2? (Service Use) {F029}
Query the hysteresis, in degrees, applied to measurement of the cascade condenser (first refrigeration stage
output), for operation of the second-stage refrigeration.
IT= / IT? {F030}
Set or query the [integral] time constant, in seconds, applied to the adjustment of the proportioning band,
governing the automatic reset ("droop" correction) of the process temperature to setpoint. Note: the IT
value is applied only to the internal temperature-control sensor. If the remote sensor is in use
(CTLREM=1), the command ITREM should be used. The proper selection of IT or ITREM is
automatically performed by the system software per the CTLREM command.
ITREM= / ITREM? {F031}
Set or query the [integral] time constant, in seconds, applied to the adjustment of the proportioning band,
governing the automatic reset ("droop" correction) of the process temperature to setpoint, when the remote
sensor is in use (Cf. IT=).
LOCK= / LOCK? {F032}
Set or query the state of the Local / Remote lockout. If set to 0, the LOCREM= command may be asserted.
If set to 1, the LOCREM= command will result in an error message.
LOCREM= / LOCREM? {F033}
Set or query the local / remote control status of the system. If set to 0, the system may only be programmed
from the [local] control panel. If set to -1, the system may only be programmed from the remote
communication link. If the system is in local control (LOCREM=0), all commands AAAAAAAA=, except
the LOCREM= command, will result in the error message "unit is in local". Queries will perform
normally. If LOCK=1 the LOCREM= message will return the error message "unit is locked in
local/remote".
LOOP2? (Service Use) (Not Yet Implemented) {F034}
Query existence of a second control loop, in dual-loop systems.