Group 1 Errors: General Parsing and Transmission Errors
E010
Transmit Buffer Overflow. Requests for information (queries) result in information being
buffered faster than it can be transmitted to the host (Baud-Rate dependent). 1KB buffer exceeded.
E011
Parse Buffer Overflow. Too many commands (>32) on a single line.
Group 2 Errors: Command Parsing Errors
E020
Undefined String. Character string not identified as valid command, in position indicated by
trailing argument.
E021
Illegal Character. Character in string not a recognized character (A-Z, a-z, 0-9, "=", "?", ".", "+",
"-", <SP> (space), <LF> (line feed), or <CR> (carriage return).
E022
Illegal Operand. Keyword string followed by illegal operation (command, program, query). E.g.,
TEMPST1 may only be queried; is a temperature which exists in the system and is measured. Attempt to
program via TEMPST1= will yield E022.
E023
Question Mark (?) Followed by Argument. Query will not accept an argument.
E024
Value Too Long. Programmed value may not exceed 8 characters.
E025
Illegal Sign Placement or Excess Decimal Points. Polarity indication (+,-) must be first character.
Only one decimal point permitted per number.
E026
Too Many Decimal Places. Too many digits supplied after decimal point.
E027
Value Out of Bounds. Programmed number exceeds limits fixed internally or by superseding
limits (e.g., FSPANH will limit USPANH).
E028
Excessive number. The previous filters will allow certain [illegal] ranges of numeric possibility to
get through. This error traps these ranges before they cause software mathematics errors (due to register
overflow, etc.).
Group 3: Security Errors
E030
Unit Not in Remote. Attempt has been made to program a value which may not be altered if
system is presently switched to "local" operation.
E031
Not a User Function. Certain program statements are reserved for factory access. Error 31
reports an attempt to program values requiring [factory] security-code access.
Group 4: Operational and Miscellaneous Errors
E040
Not Yet Implemented. Use of a command which is defined to the parser but not presently
implemented by FTS.
E041
Stop Error. System already stopped has received a "STOP" command.
E042
Start Error. System already started has received a "START" command.
E043
Not a User Function. Certain operational commands are defined to the system and reserved for
factory use. Error 43 reports use of such a command without proper security access.
E044
Local Lockout. Unit is locked in local. Removal of this lock may only be accomplished from the
system control panel before changes to system state may be made from the RS-232.