FC Series w/ EDC Control
Thermal
MNL-047-A
Division Rev.
004
FTS
34
Stone Ridge, NY USA
845.687.0071
REFRHRS? {F078}
Query running hours of system refrigeration.
REFRSW= / REFRSW? {F051}
Set or query state of refrigeration switch. 0 = off, -1 = on.
REV? {F052}
Query [embedded] software revision level in system program ROM.
RFC
Command to restore factory-determined calibration constants.
RR= / RR? {F054}
Set or query setpoint transition rate (Ramp Rate), in degrees per second.
STOPBITS? {F055}
Query the programmed number of Stop bits in the RS-232 communication definition. Values may be 1,
1.5, or 2.
SP= / SP? {F057}
Query or set the system temperature setpoint, in degrees.
SSPANH? {F058}
Query the factory-set maximum allowable setpoint, in degrees (determined by system hardware
complement).
SSPANL? {F059}
Query the factory-set minimum allowable setpoint, in degrees (determined by system hardware
complement).
START / START? {F060}
Command to turn the system on (subject to positions of PUMPSW and REFRSW) and to initiate the system
temperature control software. Query to determine whether the system is running (0 = off, -1 = on).
STATUS? (Not implemented, obsolete).
STOP
Command to turn the system off, restore displays and controls to idle.
SYSHOURS? {F063}
Query number of running hours accumulated on the system, as determined by the number of running hours
on the pump.
TEMPLI? (Service use) {F064}
Query temperature, in degrees, of suction line, for control of liquid-injection valve in refrigeration
hardware.
TEMPST1? (Service use) {F065}
Query temperature, in degrees, of the cascade condenser (heat-exchanger output of the first refrigeration
stage in a two-stage system).
TIME= / TIME? (Not yet implemented) {F066}
Set or query time of day in Real-Time Clock option.