2 Safety Information
XVIII
FOCUSED PHOTONICS INC
z
When operation is performed in a damp environment, make sure that water is kept off
the equipment.
Warning:
Non-standard and improper high voltage operations may result in fire and electric shock.
Therefore, AC cable bridging and wiring through a certain area must follow the local rules and
regulations. The personnel who perform high voltage operations should be qualified for high
voltage and AC operations.
2.2.2
Power Cable
Note:
Installation and removal of live power cable is prohibited strictly. Transient contact between
the core of power cable and conductor may generate electric arc or spark, which may lead to fire
or eye injury.
z
Before the power cable is installed or removed, the power switch must be turned off.
z
Before the power cable is connected, it must be confirmed that the power cable and label
comply with the requirements of the actual installation.
Note:
For DC power supplied equipment, please use 2.5 mm2 or 12 AWG minimum power supply
cord.
For AC power supplied equipment, please use 1.0 mm2 or 16 AWG minimum power supply
cord.
Use type H03VV-F or light PVC sheathed flexible cord according to IEC 60227.
2.2.3
Tools
Warning: