7 Operations of CEMS2000 Monitoring Software
84
FOCUSED PHOTONICS INC
Fig 7-43
Calibration Parameter
The meanings of items in this window are:
z
SO
2
calibration density: Set the calibration density of SO
2
z
NO calibration density: Set the calibration density of NO
z
Interval of calibration: Set the interval for automatic calibration between two
calibrations.
z
Calibration time: Set the duration of automatic calibration.
z
Interval of zero adjustment: Set the interval for automatic zero adjustment between two
adjustments.
z
Zero adjustment time: Set the duration of automatic adjustment.
z
Stabilizing time of zero adjustment: Set the stabilizing time of zero adjustment which is
counted from the system reaching zero point or the density is stabilized. If the
stabilizing time is passed, the system will automatically start the adjustment
You can click the “Read” button to read the parameter information of the calibration and click
the “Set” button to save the settings. Click the “
” button at the top right of the window to return
to the main window.
Note:
After the “Start adjustment / calibration” command has been sent, the system will not start
the adjustment / calibration immediately but open the relevant electro-valve first and input the
standard gas. Only after the density of standard gas is stabilized for a certain time, the system will
start the zero adjustment.