6 Operations of OMA-2000 Spectroscopic Gas Analyzer
45
FOCUSED PHOTONICS INC
Fig 6-25
Humidity Parameter Setup
¾
Current humidity: Display the humidity of the stack gas;
¾
Measurement method: Select the humidity measurement method of stack gas
display and output with optional of dry and wet;
¾
Dry: The output value after system test already includes water compensation. The
users do not need to consider the water influence;
¾
Wet: The output value after system test does not include water compensation. The
users should consider the water influence;
¾
Humidity input method: Select the input method of the humidity with measuring
input and manual input.
When the measuring input is selected, the system measures the humidity from internal
humidity sensor while when the manual input is selected the user can manually input the humidity
parameter. The setup is the same as the setup for temperature.
z
Flow rate: Used for checking the current of the gas flow rate and setting the offset of
zero point and smoothness times, as shown in Fig 6-26.