Operation Manual EMGZ 309.EIP
20.12.2018
36
2.
The System Force [SysForce] determines the measuring capability of your
measuring roller. E.g. if you use two 500 N sensors in your measuring roller,
enter 1000N (see chapter 8.3 “Descriptions of Parameter”, [SysForce]).
3.
Enter the [SysForce] in the corresponding box by overwriting the default or
previous value.
4.
Enter the force corresponding to your calibration weight in the box “Weight”
5.
Click on the box “Calibrate Gain” to start the calibration procedure.
6.
The Gain value will be calculated, indicated in the box “Gain” and stored in the
parameter [Gain]. The calibration procedure has been completed successfully.
B)
For Devise Mode Volt Gauge
1.
Load a rope with a defined weight corresponding to your calibration force on
the roller. The roller configuration must correspond to the real configuration in
the machine (wrap angle, distance of the rollers etc.).
2.
Enter the force corresponding to your calibration weight in Volt (V) in the field
“Weight”. If this value is not known, the calibration method with the FMS-
Calculator can be used to determine it. (see chapter 4.9 „Analytic Calibration
Method“).