DTX Series CableAnalyzer
Technical Reference Handbook
8-58
To use the power meter in MONITOR mode:
Note
If you need to save the power reading, use the
power meter function in the SINGLE TEST mode.
1
Clean the tester’s input port, the test reference cord
connectors, and the source connector.
2
Use the test reference cord to connect the source to the
tester's input port, as shown in Figure 8-30.
3
Turn on the source.
4
Turn the rotary switch to MONITOR; then select Power
Meter.
5
Press
P
.
6
Select the appropriate wavelength; then press
P
. The
power meter in MONITOR mode runs continuously
until you press
I
.
Figure 8-31 describes the power meter screens for SINGLE
TEST and MONITOR modes.
amd126f.eps
Figure 8-30. Connections for Monitoring Optical Power (MONITOR mode)