DTX Series CableAnalyzer
Technical Reference Handbook
7-34
Using the Measurement Cursor
You can use the cursor on the OTDR trace to measure
distances and losses along the cabling.
To make measurements with the cursor:
1
Press
K
Change to Cursor if necessary to set the
arrow keys to cursor mode.
2
Move the cursor to a starting point for a measurement.
Use
BC
to move the cursor left or right or
AD
to move the cursor to the previous or
next event.
3
Press
H
to mark the starting point for the
measurement.
4
Use
BCAD
to move the cursor to the
ending point for the measurement.
The following readouts appear on the screen:
•
Δ
m or
Δ
ft: The distance between the cursor and
the mark.
•
Δ
dB: The difference in the power levels (in
decibels) between the cursor and the mark.
5
To clear the mark press
H
.
Summary of Contents for DTX-1800