PERFORMANCE TEST
4 - 17
Test set up:
Settings/procedure:
1
-
Apply a 1 kHz sine wave signal with an amplitude of 0.8 V (pp) to input
CH1.
2
-
Press the AUTOSET key and set CH1 to 0.1V/div.
3
-
Adjust the generator so that the displayed amplitude is exactly 8 divisions.
Requirements
-
Turn the CH1 POS control fully clockwise and counterclockwise and verify that
top and bottom of the 8 divisions signal can be positioned outside the
graticule.
Repeat the above settings and procedure for CH2, CH3, and CH4.
PM3370B/80B/90B: check CH2.
4.3.15
Vertical deflection; crosstalk between channels at 60/100/200 MHz
At higher frequencies there exists some crosstalk between any two channels. In
the following test, crosstalk is verified at a high frequency.
Test equipment:
Fluke 5500A mode: levsine (Alternative constant amplitude sine wave generator
SG 503).