PERFORMANCE TEST
4 - 55
Settings/procedure:
-
Use a power splitter and two coaxial 50
Ω
cables to apply the generator output
signal to input CH1 and the rear panel input ‘EXT TRIG’.
-
Use an external 50
Ω
termination at the end of each cable. When available,
use the internal termination of input CH1 (if active, the text ‘LZ’ is visible in the
lower part of the viewing area).
-
Adjust the generator output signal for a 50 kHz sine-wave of 200 mV (pp
output amplitude into 50
Ω)
.
-
Press the green AUTOSET key.
-
Press the TRIGGER menu key to display the TRIGGER MAIN TB menu.
Select with softkey ‘ch1, extern, line’ the ‘extern’ trigger source.
-
Put CH1 in 50 mV/div and adjust the generator output voltage to a vertical
display of 2 divisions.
-
Slowly increase the frequency of the generator signal from 50 kHz to 5 MHz.
Requirement:
Check that the signal diplayed via CH1 stays well triggered across the indicated
frequency range (the ‘ARM’D LED must stay dimmed).
Settings/procedure:
-
Adjust the generator output voltage to a vertical display of 4 divisions.
-
Slowly increase the frequency of the generator signal from 5 MHz to 10 MHz.
Requirements:
Check that the signal diplayed via CH1 stays well triggered across the indicated
frequency range (the ‘ARM’D LED must stay dimmed).