Table D-1. Computer Control Commands (cont.)
Description Command
Returned
String
Perform Leak Test at
XXXmmHg
[LEAK,XXX]
50<=XXX<=400
Returns ACK
then Leak Rate.
Returns Leak
Rate after 60
seconds,
followed by CR-
LF, or NAK if
XXX out of
range•
Perform Pressure Source at
XXXmmHg
[PSOURCE,XXX]
50<=XXX<=400
ACK or NAK if
XXX out of
range
[BLEED_SYSTEM]
ACK
Performs Relief Valve Test
[RELIEF,XXX]
50<=XXX<=400
Returns ACK
then Peak
Pressure,
followed by CR-
LF, or NAK if
XXX out of
range
Pressure Gauge
[GAUGE] Returns
ACK
then pressure
port value,
followed by CR-
LF