BP Pump 2
Operators Manual
1-6
•
Always connect the system power cord directly to a
three-prong receptacle with a functional ground. Never
use a two-prong plug adapter to connect primary
power to the Tester, thereby disconnecting the utility
ground.
•
Disconnect the Tester from the power source before
changing the supply voltage. The Tester operates at a
range of 100 to 240 volts.
W
Caution
To avoid
damage to the Tester or adverse affects on its
performance, follow these guidelines
:
•
Allow only qualified technical personnel to service the
Tester.
•
Do not expose the system to temperature extremes.
Ambient temperatures should remain between 15° C
and 40° C. System performance may be adversely
affected if temperatures fluctuate above or below this
range.
•
Clean the Tester only by gently wiping down with a
clean, lint-free cloth dampened with a mild detergent
solution. Do not immerse the unit.
•
Do not apply pressures greater than 400 mmHg (53
kPa) to the pressure port.