Operation
Simulations
3
3-19
Neonate
The Neonate simulations are provided to test the ability of the NIBP monitors to
detect blood pressure on neonatal patients.
To perform Neonate simulations, do the following:
1.
Access these simulations by pressing the
8 NEONATE
key. The
Neonate
Simulation
screen appears.
N
eonate Sim
u
lation
N
eonate # 1
C
u
ff: Internal
N
eonate
OPTIO
N
S
CUFF
ready ....
35/
15(22)
120 BPM 0.11 cc
fas12.eps
2.
Press the
OPTIONS
soft key to scroll through the simulation choices.
Parameters for Neonate simulations are shown in Table 3-6.
Table 3-6. Neonate Simulations
Simulation
Number
Blood Pressure
(mmHg) (MAP)
Heart Rate (bpm)
Pulse volume
(cc)
1 35/15
(22)
120 0.11
2 60/30
(40)
120 0.10
3 80/50
(60)
120 0.10
4 100/70
(80)
120
0.10
3.
Press the
CUFF
soft key to select External or Internal Neonate.
4.
Press
Start
on the NIBP unit.
5.
Compare values on the NIBP unit with those on the Tester.
6.
Press
HOME
to return to the
Main
menu.
Summary of Contents for BP Pump 2 Operators
Page 6: ......
Page 10: ...BP Pump 2 Operators Manual iv...
Page 12: ...BP Pump 2 Operators Manual vi...
Page 14: ...BP Pump 2 Operators Manual 1 2...
Page 26: ...BP Pump 2 Operators Manual 2 2...
Page 38: ...BP Pump 2 Operators Manual 3 2...
Page 68: ...BP Pump 2 Operators Manual...
Page 74: ...BP Pump 2 Operators Manual B 4...
Page 82: ...BP Pump 2 Operators Manual D 6...