Operation
Pressure
Tests
3
3-7
Conversion Factors
Conversion factors for the Tester are shown in Table 3-1.
Table 3-1. Conversion Factors
Units mmHg
PSI_PER_MMHG 0.019337
CMH2O_PER_MMHG 1.3595
INH2O_PER_MMHG 0.53525
KPA_PER_MMHG 0.13332
Initializing Tests and Simulations
You can access functions of the Tester in two ways:
•
Exclusive use of soft keys, the row of up arrow keys just below the display
•
Combined use of the number keys and soft keys
For example, you may access a Pressure Source test as follows:
Soft keys only:
PRESSURE TESTS
C
STATIC PRESSURE
C
GAUGE
C
SOURCE
Combined:
3 STATIC PRESSURE
C
GAUGE
C
SOURCE
Either method is acceptable, and it is assumed that you will develop your own,
most effective way of using the keys.
Error Messages
If you encounter any repeatable error messages, contact your Fluke Biomedical
Service Center.
Pressure Tests
The following tests assess the integrity and accuracy of the Tester, as well as the
instrument under test.