BP Pump 2
Operators Manual
3-4
shape of the pulses are controlled by the microprocessor driving the step motor.
The home position of the piston is detected by an optical interrupter. If the
optional high-accuracy pressure sensor is installed, it also measures the pressure in
the main manifold and is also controlled by the microprocessor.
Configurations for Devices Under Test (DUT)
Connect the Tester to the NIBPM unit with the tubing and fittings (part number
2391882) supplied with the Tester in the desired configurations, as shown in
Figures 3-2 through 3-6.
Basically, configurations involve connecting the NIBPM unit to the Tester directly
or connecting the unit, a cuff, and the Tester, using a “T” connector. When
connected properly, the tubing forms a closed system with the components.
BP Pump 2
NON-INVASIVE BLOOD PRESSURE MONITOR ANALYZER
PRESSURE
LEAK
1
PRESSURE
RELIEF
TESTS AND SIMULATIONS
2
STATIC
PRESSURE
3
STANDARD BP
NEONATE
WRIST
PRESSURE PORT
USER DEFINED
ARRHYTHMIAS
4
PATIENT
CONDITIONS
5
6
9
8
RESPIRITORY
ARTIFACT
7
0
ENT
120
93
8
0
8
0
N
IBP Monitor
Single Hose: Internal C
u
ff
fas16.eps
Figure 3-2.Connecting Tester to Single-hose NIBP Monitor
(Int Cuff)
Summary of Contents for BP Pump 2 Operators
Page 6: ......
Page 10: ...BP Pump 2 Operators Manual iv...
Page 12: ...BP Pump 2 Operators Manual vi...
Page 14: ...BP Pump 2 Operators Manual 1 2...
Page 26: ...BP Pump 2 Operators Manual 2 2...
Page 38: ...BP Pump 2 Operators Manual 3 2...
Page 68: ...BP Pump 2 Operators Manual...
Page 74: ...BP Pump 2 Operators Manual B 4...
Page 82: ...BP Pump 2 Operators Manual D 6...