8845A Accuracy
Accuracy is given as
±
(% measu % of range)
Range
Frequency
(Hz)
24 Hour
(23
±
1
°
C)
90 Days
(23
±
5
°
C)
1 Year
(23
±
5
°
C)
Temperature
Coefficient/
°
C
Outside 18 to 28
°
C
3 – 5 Hz
1.0 + 0.03
1.0 + 0.04
1.0 + 0.04
0.10 + 0.004
5 – 10 Hz
0.35 + 0.03
0.35 + 0.04
0.35 + 0.04
0.035 + 0.004
10 Hz – 20 kHz
0.04 + 0.03
0.05 + 0.04
0.06 + 0.04
0.005 + 0.004
20 – 50 kHz
0.1 + 0.05
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
100 mV
100 – 300 kHz
[1]
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
3 – 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
1 V
100 – 300 kHz
[1]
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
3 – 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
10 V
100 – 300 kHz
[1]
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
3 – 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
100 V
100 – 300 kHz
[1]
4.0 + 0.50
4.0 + 0.50
4.0 + 0.50
0.2 + 0.02
3 –- 5 Hz
1.0 + 0.02
1.0 + 0.03
1.0 + 0.03
0.1 + 0.003
5 – 10 Hz
0.35 + 0.02
0.35 + 0.03
0.35 + 0.03
0.035 + 0.003
10 Hz – 20 kHz
0.04 + 0.02
0.05 + 0.03
0.06 + 0.03
0.005 + 0.003
20 – 50 kHz
0.1 + 0.04
0.11 + 0.05
0.12 + 0.05
0.011 + 0.005
50 – 100 kHz
[2]
0.55 + 0.08
0.6 + 0.08
0.6 + 0.08
0.06 + 0.008
750 V
100 – 300 kHz
[1] [2]
4.0 + 0.5
4.0 + 0.5
4.0 + 0.5
0.2 + 0.02
[1] Typically 30 % reading error at 1 MHz
[2] 750 Volt range is limited to 8 X 10
7
volt-Hertz
Additional Low Frequency Errors
Error is stated as % of reading.
AC Filter
Frequency
3 HZ (slow)
20 HZ (medium)
200 HZ (fast)
10 – 20 Hz
0
0.25
–
20 – 40 Hz
0
0.02
–
40 – 100 Hz
0
0.01
0.55
100 – 200 Hz
0
0
0.2
200 Hz – 1 kHz
0
0
0.02
>1
kHz
0
0
0
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...