Preparing the Meter for Operation
Connecting to Line Power
2
2-7
Connecting to Line Power
XW
Warning
To avoid shock hazard, connect the factory supplied three-
conductor line power cord to a properly grounded power outlet.
Do not use a two-conductor adapter or extension cord; this will
break the protective ground connection. If a two-conductor
power cord must be used, a protective grounding wire must be
connected between the ground terminal and earth ground
before connecting the power cord or operating the instrument.
First verify that the line voltage selection is set to the correct position, and then that the
correct fuse for that line voltage is installed. Connect the Meter to a properly grounded
three-prong outlet.
Table 2-2. Line Power Cord Types Available from Fluke
Type
Voltage/Current
Fluke Model Number
North America
120V/15A
LC-1
North America
240V/15A
LC-2
Universal Euro
220V/16A
LC-3
United Kingdom
240V/13A
LC-4
Switzerland 220V/10A LC-5
Australia 240V/10A LC-6
South Africa
240V/5A
LC-7
LC-1
LC-2
LC-3
LC-4
LC-5
LC-6
LC-7
alh3.eps
Figure 2-3. Line Power Cords Types Available from Fluke
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...