8845A/8846A
Users Manual
4-16
None
Checking Diodes
The diode function sends a current through a semiconductor junction while the Meter
measures the voltage drop across the junction (or junctions). Measurements are displayed
in the 10 V range at a relatively fast measurement rate. "
OPEN
" is displayed for voltages
10 % above the compliance voltage setting. Typical junction voltage drop on good
junctions are 0.3 to 0.8 volts. If enabled, the beeper will make a short beep when a good
junction is detected. Shorted diodes will indicate a significantly lower voltage.
With the higher compliance voltage (up to 10 V), the Meter’s diode test function can test
zeners up to 10 volts, diode stacks and LEDs. The selectable current and maximum
voltage allows you to tailor the diode test to the expected voltage for the junction being
tested.
To check a diode:
1.
Press
S
twice on the 8846A or
D
once on the 8845A. An example of the diode
test display is shown below.
0.5994
F1
F2
F3
F4
F5
+
caw13f.eps
2.
Select the test voltage and current appropriate for the diode being tested by pressing
the associated soft keys.
3.
Connect the test leads as shown in Figure 4-8.
2
INPUT
SENSE
HI
LO
FRONT
REAR
10A
1V
00
mA
4W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
600V CAT II
4
caw024.eps
Figure 4-8. Diode Testing Connections
Function modifiers:
None
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...