8845A/8846A
Users Manual
3-22
Removing the Power-up Configuration
To remove the Meter’s stored power-up configuration:
4.
Press
M
.
5.
Press the
STORE CONFIG
softkey.
6.
Press the
REMOVE POWER-UP
softkey.
After removing the power-up configuration, power the Meter down and back up using the
rear-panel power switch to restore factory default settings.
Recalling a Meter Configuration
To recall a configuration from internal memory:
1.
Press
M
.
2.
Press the
RECALL CONFIG
soft key.
3.
Press the
RECALL INT MEM
soft key.
4.
Press the soft key labeled with the memory location (
CONFIGA
through
CONFIGE
).
To recall a configuration from external memory (8846A Only):
1.
Press
M
.
2.
Press the
RECALL CONFIG
soft key.
3.
Press the
RECALL USB
soft key.
The Meter labels the first four soft keys with the first four memory locations
CONFIG01
through
CONFIG04
. The fifth soft key is labeled
MORE
to allow
access to all 100 memory locations.
4.
To recall the Meter configuration from one of the first four memory locations, press
the appropriately labeled soft key. If you want a memory location other than the first
four locations, press the
MORE
soft key.
The display indicates the last memory location containing a Meter configuration. If
all configuration memory locations are filled, the Meter will always indicate memory
location 10.
5.
To select the memory location, press either
<--
or
-->
to select a specific digit.
With the digit selected, press the soft key labeled
--
to decrement the digit or
++
to
increment the digit.
6.
Once the chosen memory location is set, press the
ENTER
soft key to recall that
Meter configuration.
Note
Incompatible configurations (from incompatible configuration versions)
will not be loaded, but will generate error +229 "Incompatible
measurement configuration not loaded".
Managing Memory
The Meter provides a method of clearing internal memory and displaying the status of
external memory (8846A Only). Per Department of Defense requirements, the Meter
allows the erasure of Meter configuration and data files from external USB memory
modules. Other files are not erased from the modules.
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...