Introduction and Specifications
General Specifications
1
1-9
Table 1-5. Accessories (cont)
Model/Fluke PN
Description
1259340
Agreement, Calibration, Accredited
2441827
Agreement, Calibration, primary standards lab
1540600
Agreement, Calibration, artifact
General Specifications
Power
Voltage
100 V Setting ...................................................... 90 V to 110 V
120 V Setting ...................................................... 108 V to 132 V
220 V Setting ...................................................... 198 V to 242 V
240 V Setting ...................................................... 216 V to 264 V
Frequency ............................................................... 47 Hz to 440 Hz. Automatically sensed at power-on.
Power Consumption................................................ 28 VA peak (12 Watt average)
Dimensions
Height...................................................................... 88 mm (3.46 in.)
Width....................................................................... 217 mm (8.56 in.)
Depth ...................................................................... 297 mm (11.7 in.)
Weight..................................................................... 3.6 kg (8.0 lb)
Shipping Weight...................................................... 5.0 kg (11.0 lb)
Display
Vacuum Fluorescent Display, dot matrix
Environment
Temperature
Operating ............................................................ 0
°
C to 55
°
C
Storage ............................................................... -40
°
C to 70
°
C
Warm Up............................................................. 1 hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating ............................................................ 0
°
C to 28
°
C <90 %
28
°
C to 40
°
C <80 %
40
°
C to 55
°
C <50 %
Storage ............................................................... -40
°
C to 70
°
C <95 %
Altitude
Operating ............................................................ 2,000 Meters
Storage ............................................................... 12,000 Meters
Vibration and Shock................................................ Complies with Mil-T-28800F Type III, Class 5 (Sine only)
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V
EMC
Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This Meter has
shown susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz.
Triggering
Samples per Trigger ........................................... 1 to 50,000
Summary of Contents for 8845A
Page 6: ...8845A 8846A Users Manual iv...
Page 8: ...8845A 8846A Users Manual vi...
Page 10: ...8845A 8846A Users Manual viii...
Page 12: ...8845A 8846A Users Manual 1 2 Diode Test 1 22 Measurement Rates IEEE488 4 1 22...
Page 34: ...8845A 8846A Users Manual 2 2...
Page 42: ...8845A 8846A Users Manual 2 10...
Page 70: ...8845A 8846A Users Manual 4 2...
Page 88: ...8845A 8846A Users Manual...
Page 90: ...8845A 8846A Users Manual A 2...
Page 96: ...8845A 8846A Users Manual C 2...
Page 102: ...8845A 8846A Users Manual 4...