5560A/5550A/5540A
Operators Manual
56
Verification Procedure
1. Set the DUT function to the position called for in Table
2. Apply the indicated Calibrator output to the DUT for each test.
3. To verify the performance of the DUT, make sure that the DUT reads within the limits shown in the
Meter Response column.
Table 15. Performance Test - 77 Series IV DMM
Steps
Test
(Switch Position)
Calibrator Output
Meter Response
Lower Limit
Upper Limit
1
e
Ohms
[2]
500
Ω
497.3
Ω
502.7
Ω
2
5 k
Ω
4.974 k
Ω
5.026 k
Ω
3
50 k
Ω
49.74 k
Ω
50.26 k
Ω
4
5 M
Ω
4.974 M
Ω
5.026 M
Ω
5
10 M
Ω
9.79 M
Ω
10.21 M
Ω
6
40 M
Ω
39.19 M
Ω
40.81 M
Ω
7
R
Continuity
25
Ω
Beeper On
8
250
Ω
Beeper Off
9
K
AC Volts
50 mV 45 Hz
0.047 V ac
0.053 V ac
10
5 V 45 Hz
4.898 V ac
5.102 V ac
11
5 V 1 kHz
4.898 V ac
5.102 V ac
12
50 V 45 Hz
48.98 V ac
51.02 V ac
13
50 V 1 kHz
48.98 V ac
51.02 V ac
14
500 V 45 Hz
489.8 V ac
510.2 V ac
15
500 V 1 kHz
489.8 V ac
510.2 V ac
16
1000 V 45 Hz
978 V ac
1022 V ac
17
1000 V 1 kHz
978 V ac
1022 V ac
18
K
Hz
AC Volts Frequency
[1]
5 V 99 Hz
98.89 Hz
99.11 Hz
19
5 V 900 Hz
899 Hz
901 Hz
21
5 V 50 kHz
49.94 kHz
50.06 Hz
22
L
DC Volts
5 V
4.984 V dc
5.016 V dc
23
50 V
49.84 V dc
50.16 V dc
24
300 V
299.0 V dc
301.0 V dc
25
1000 V
996 V dc
1004 V dc
26
-1000 V
-1004 V dc
-996 V dc