Calibrator
Sample Applications
55
Calibrate a 77 Series IV DMM
Note
This is an example procedure. The 77 Series IV Calibration Information Manual contains the
authoritative verification and adjustment procedures for that product.
This example uses the Fluke 55XXA/DMMCAL cable assembly and the Product to verify a Fluke 77
Series IV DMM.
55XXA/DMMCAL Cable Assembly
XW
Warning
To prevent electric shock, fire, or personal injury, make sure the product is in Standby before
making any connection between the product and tester.
Use the 55XXA/DMMCAL cable assembly to connect the DUT to the Calibrator. See Figure
.
1. Connect the stackable double banana plug labeled DMM SENSE to the DUT inputs:
z
and
COM
.
2. Stack the double banana plug labeled DMM INPUT into the DMM SENSE plug from step 1.
3. Connect the single banana plug labeled LO I INPUT to the DUT
400mA
input.
4. Connect the single banana plug labeled HI I INPUT to the DUT
10A
input.
Figure 13. 55XXA/DMMCAL Cable Connection to a 77 IV Series DMM