5522A
Operators Manual
9-14
1.
Connect the 5522A to Channel 1 on the oscilloscope. Select 50
Ω
impedance or use a
50
Ω
termination directly at the oscilloscope input. Verify that the
O
key is lit,
indicating that the signal is connected.
2.
Alter the voltage setting for the signal so it matches the amplitude value
recommended by your oscilloscope manufacturer for calibrating the edge response.
The default setting is 25 mV @ 1 MHz.
For example, on a Fluke PM3392A oscilloscope, start with a signal of 1 V @ 1 MHz.
3.
Adjust the scale on your oscilloscope to achieve a good picture of the edge. For
example, on a Fluke PM3392A oscilloscope with a 1 V @ 1 MHz signal, use
200 mV/div.
4.
Adjust the time base on your oscilloscope to the fastest position available (20.0 or
50.0 ns/div).
Pulse aberrations
gl007i.eps
5.
Verify that your oscilloscope exhibits the proper rise time and pulse aberration
characteristics.
6.
Remove the input signal by pressing
Y
.
Pulse Response Calibration with a Tunnel Diode Pulser
You can use the calibrator to drive a tunnel diode pulser (Fluke Part Number 606522, or
Tektronix 067-0681-01, or equivalent), allowing you to check for pulse edge rise times as
fast as 125 ps.
The calibrator sources a maximum pulser drive signal of 100 V p-p at 100 kHz. The
recommended (and default) output setting is 80 V p-p at 100 kHz.
Perform the following procedure to use a tunnel diode pulser:
1.
Connect the Calibrator, tunnel diode pulser, and oscilloscope as shown in Figure 9-2.
2.
With the SC600 Option in EDGE mode, press the TDPULSE softkey to “on”.
3.
Press
O
.
4.
Rotate the control on the pulser box to the minimum setting necessary to trigger a
reading.
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...