5522A
Operators Manual
6-38
<interface>,
TERM
(terminal), COMP (computer)
<flow
control>,
XON
(xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>,
DBIT7 (7 bits) or
DBIT8
(8 bits)
<number
stop
bits>,
SBIT1
(1 bit) or SBIT2 (2 bits)
<parity>,
PNONE
(none), PODD (odd),PEVEN
(even)
<end of line char.>
CR (carriage return), LF (line feed),
CRLF
(carriage return/line feed)
Example:
SP_SET 9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
Set the parameters for the rear panel SERIAL 1 FROM HOST serial port to the factory
default values.
SP_SET?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Host Serial Port Set query) Returns the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port. (To return the parameters for the rear panel
SERIAL 2 TO UUT serial port, see the
UUT_SET?
command.) The factory default
values are shown below in
bold
type. (To return to the factory defaults, see the
FORMAT
SETUP
command.)
Responses: <baud rate value>,
300, 600, 1200, 2400, 4800,
9600
<interface>,
TERM
(terminal), COMP (computer)
<flow
control>,
XON
(xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>,
DBIT7 (7 bits) or
DBIT8
(8 bits)
<number
stop
bits>,
SBIT1
(1 bit) or SBIT2 (2 bits)
<parity>,
PNONE
(none), PODD (odd),PEVEN
(even)
<end of line char.>
CR (carriage return), LF (line feed),
CRLF
(carriage return/line feed)
Example:
SP_SET?
returns
9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
Returns the parameters for the rear panel SERIAL 1 FROM HOST serial port, as shown,
when set to the factory default values.
SPLSTR
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Serial Poll String command) Sets the Serial Poll String (string up to 40 characters)
which is saved in the Calibrator non-volatile memory. (While saving configuration data
in the non-volatile memory, a period of about 2 seconds, the Calibrator does not respond
to remote commands.) The SPLSTR is sent to the host over the serial interface when a ^P
(<cntl> P) character is sent. The default format is:
SPL: %02x %02x %04x %04x
where the term %02x (8 bits) means print the value in hexadecimal with exactly 2 hex
digits, and %04x (16 bits) means print the value in hexadecimal with exactly 4 hex digits.
The string representations are:
SPL: (STB) (ESR) (ISCR0) (ISCR1)
See the commands, respectively,
*STB?
,
*ESR?
,
ISCR0?
, and
ISCR1?
. A typical
string in the default format sent to the host is:
SPL: 44 00 0000 1000
. This
command is for format. For values instead of format, enter a ^P (<cntl> p) character.
Also see the
SRQSTR
command.
Parameter:
“<string>\n”
(
\n
represents the NEWLINE character, hex 0A)
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...