5522A
Operators Manual
5-8
Chapter 9, “Accessories.”) A typical connection is shown in Figure 5-3.
UUT
IEEE-4
88
Port
IEEE-4
88
Port
5522A Cali
b
rator
Controller
IEEE-4
88
Ca
b
le
gjh043.eps
Figure 5-3. Testing the IEEE-488 Port
Complete the following procedure to test IEEE-488 operation using Win32 Interactive
Control.
1.
Complete the “IEEE-488 Port Setup Procedure” earlier in this chapter to set up the
Calibrator for GPIB operation. Note the GPIB Address Port (default is 4).
2.
Connect the PC and Calibrator IEEE-488 ports using a standard IEEE-488 cable.
(See Chapter 9, “Accessories,” for IEEE-488 cables available from Fluke.)
3.
From the programs menu, select "NI-488.2M software for...(your operating system)".
4.
From the NI488.2M software menu, select "Win32 interactive control".
5.
A DOS window opens with a prompt as shown here:
[
6.
At the prompt type the following line to activate the IEEE interface card:
<ibdev 0 4 0 10 1 0>
The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
7.
The prompt reads <ud0:>. From this prompt type <ibwrt "remote"> then press the
ENTER (or RETURN) key.
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...