Example:
ZCOMP?
returns
NONE
Returns
NONE
when no impedance compensation is applied to the resistance, capacitance
or RTD output.
ZERO_MEAS
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Zero Offset for Pressure Measurement command) Zeros the pressure module or sets the
zero offset for capacitance measurement using the -SC600. For pressure measurments, if
the pressure module is an absolute module, the reference parameter must be supplied
along with optional units as the second argument.
Parameter:
1. (boolean) ON
(boolean) OFF
2. Reference value for absolute pressure modules
Example:
ZERO_MEAS ON
Sets the zero offset to the present measurement value.
Example:
ZERO_MEAS ON,14.7
Sets the zero offset to 14.7 for an absolute pressure module.
ZERO_MEAS?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Zero Offset for Pressure Measurement query) Returns the zero offset for the pressure
module or capacitance measurement using the -SC600.
Parameter:
(optional)
units of returned value
Responses: 1. (character) OFF
(no zero in effect)
(character) ON
(zero in effect)
2. (float) offset value
3. (character) units
(F, PSI, MHG, INHG, INH2O, FTH2O, MH2O,
BAR,
PAL,
G/CM2,
INH2O60F)
Example:
ZERO_MEAS?
returns
ON,-3.66E-02,PSI
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...