Remote Commands
Commands
6
6-27
If you change the frequency of an ac function and the harmonic output is not explicitly
set at the same time with the
HARMONIC
command, the harmonic will be set to 1.
Use multipliers e.g., k, M,
μ
with the
OUT
command, as desired.
Parameters: <value> V
Volts dc or update volts ac
<value> DBM
Volts ac dBm update
<value> V, <value> Hz
Volts ac or volts dc with 0 Hz
<value> DBM, <value> Hz
Volts ac in dBm
<value> A
Current dc or update current ac
<value> A, <value> Hz
Current ac
<value>
OHM
Resistance
<value>
F
Capacitance
<value>
CEL
Temperature
(Celsius)
<value>
FAR
Temperature
(Fahrenheit)
<value> HZ
Update frequency
<value> V, <value> A
Power dc or update power ac
<value> V, <value> A, <value> HZ
Power ac
<value> V, <value> V
Dual volts dc or update dual ac
<value> V, <value> V, <value> HZ
Dual volts ac in volts
<value>
For single output, changes
amplitude keeping unit and
frequency the same.
Examples:
OUT 15.2 V
(volts; 15.2 V @ same frequency)
OUT 20 DBM
(volts; 20 dBm @ same frequency)
OUT 10 V, 60 Hz
(volts ac; 10 V @ 60 Hz)
OUT 10 DBM, 50 HZ
(volts ac; 10 dBm @ 50 Hz)
OUT 1.2 MA
(current; 1.2 mA @ same frequency)
OUT 1 A, 400 HZ
(current ac; 1 A @ 400 Hz)
OUT 1 KOHM
(ohms; 1 k
Ω
)
OUT 1 UF
(capacitance;
1
μ
F)
OUT 100 CEL
(temperature;
100
°
C)
OUT 32 FAR
(temperature;
32
°
F)
OUT 60 HZ
(frequency update; 60 Hz)
OUT 10 V, 1 A
(power; 10 watts @ same frequency)
OUT 1 V, 1 A, 60 HZ
(power ac; 1 watts @ 60 Hz)
OUT 1 V, 2 V
(dual volts; 1 V, 2 V @ same freq.)
OUT 10 MV, 20 MV, 60 HZ
(dual volts; .01 V, .02 V @ 60 Hz)
Each example shows a value and unit, e.g., –15.2 V. If a value is entered without a unit,
the value of the existing output is changed, when logically allowed.
OUT?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Output query) Returns the output amplitudes and frequency of the Calibrator.
Multipliers (e.g., K or M) are not used in the response.
Parameters:
V
(optional for ac voltage and TC modes)
DBM
(optional for ac voltage modes)
CEL
(optional for RTD and TC modes, Celsius)
FAR
(optional for RTD and TC modes, Fahrenheit)
OHM
(optional for RTD modes, ohms)
Response:
<primary amplitude value>,<primary units>,
<secondary amplitude value>,<secondary units>,
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...