AC Current Specifications, LCOMP OFF (continued)
Range,
Composite
Waveform
Harmonic
Frequency
Harmonic
Amplitude
Range
(% of
Fundamental)
[
1]
Harmonic Amplitude
Uncertainty
(% of
Funda A)
Harmonic Phase
Uncertainty
(Relative to
Fundamental)
[
2
]
Absolute RMS
Uncertainty of
Composite
Waveform
(% RMS + A)
15 to 45 Hz
0.1 to 100 %
0.1 % + 1 mA
0.5
°
45 to 900 Hz
0.1 to 100 %
0.1 % + 1 mA
0.6
°
900 Hz to 2 kHz
0.1 to 20 %
0.1 % + 1 mA
1
°
0.33 to
2.999 A
2 to 5 kHz
0.1 to 20 %
0.2 % + 1.3 mA
2
°
0.2 % + 1 mA
15 to 45 Hz
0.1 to 100 %
0.1 % + 10 mA
0.5
°
45 to 900 Hz
0.1 to 100 %
0.1 % + 10 mA
0.6
°
900 Hz to 2 kHz
0.1 to 20 %
0.1 % + 10 mA
1
°
3 to 20.5 A
2 to 5 kHz
0.1 to 20 %
0.2 % + 10 mA
3
°
0.2 % + 10 mA
[1]
All frequencies can have harmonics up to 100 % of the fundamental; uncertainties are not specified unless otherwise indicated.
[2]
For harmonics that are <1 % of the Fundamental, phase uncertainty is typical.
AC Current Specifications, LCOMP ON
*
Range,
Composite
Waveform
Harmonic
Frequency
Harmonic
Amplitude
Range (% of
Fundamental)
[1]
Harmonic Amplitude
Uncertainty
(% of
Funda A)
Harmonic Phase
Uncertainty
(Relative to
Fundamental)
[2]
Absolute RMS
Uncertainty of
Composite
Waveform
(% RMS + A)
15 to 65 Hz
0.1 to 30 %
0.5 % + 0.1
μ
A 0.5
°
29 to
329.99
μ
A
65 to 900 Hz
0.1 to 30 %
1.0 % + 0.1
μ
A 2
°
0.5 % + 1
μ
A
15 to 65 Hz
0.1 to 30 %
0.5 % + 1
μ
A 0.5
°
0.33 to
3.2999 mA
65 to 900 Hz
0.1 to 30 %
1.0 % + 1
μ
A 1
°
0.5 % + 1
μ
A
15 to 65 Hz
0.1 to 30 %
0.4 % + 10
μ
A 0.5
°
3.3 to
32.999 mA
65 to 900 Hz
0.1 to 30 %
0.6 % + 10
μ
A 1
°
0.5 % + 10
μ
A
15 to 65 Hz
0.1 to 30 %
0.4 % + 100
μ
A 0.5
°
33 to
329.9 mA
65 to 900 Hz
0.1 to 30 %
0.6 % + 100
μ
A 1
°
0.5 % + 100
μ
A
15 to 65 Hz
0.1 to 30 %
0.5 % + 1 mA
0.75
°
0.33 to
2.999 A
65 to 440 Hz
0.1 to 30 %
1.0 % + 1 mA
1
°
0.5 % + 1 mA
15 to 65 Hz
0.1 to 30 %
0.5 % + 10 mA
0.75
°
3 to 20.5 A
65 to 440 Hz
0.1 to 30 %
1.0 % + 10 mA
1
°
0.75 % + 10 mA
* LCOMP ON is used to drive inductive loads like the 5500A/COIL and current clamps.
[1]
All frequencies can have harmonics up to 100 % of the fundamental; uncertainties are not
specified unless otherwise indicated.
[2]
For harmonics that are <1 % of the Fundamental, phase uncertainty is typical.
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...