5522A
Operators Manual
7-20
Table 7-15. Verification Tests for Frequency
Range, Normal
Output, V
Output, Normal,
V
Frequency Lower
Limit
[1]
Upper
Limit
[1]
119.00 Hz
118.99970 Hz
119.00030 Hz
120.0 Hz
119.99970 Hz
120.00031 Hz
1000.0 Hz
999.9975 Hz
1000.0025 Hz
3.29999 3.00000
100.00 kHz
99,999.75 Hz
100,000.25 Hz
[1]
Frequency accuracy is specified for 1 year.