5522A
Operators Manual
2-4
How to Select Line Voltage
The calibrator arrives from the factory configured for the line voltage normally
appropriate for the country of purchase, or as specified at the time of your purchase order.
You can operate the Calibrator from one of four line voltage settings: 100 V, 120 V, 200
V, and 240 V (47 Hz to 63 Hz). To check the line voltage setting, note the voltage setting
visible through the window in the power line fuse compartment cover (Figure 2-1). The
allowed line voltage variation is 10% above or below the line voltage setting.
To change the line voltage setting, complete the following procedure:
1.
Disconnect line power
.
2.
Open the fuse compartment by inserting a screwdriver blade in the tab located at the
left side of the compartment and gently pry until it can be removed with the fingers.
3.
Remove the line voltage selector assembly by gripping the line voltage indicator tab
with pliers and pulling it straight out of its connector.
4.
Rotate the line voltage selector assembly to the desired voltage and reinsert.
5.
Verify the appropriate fuse for the selected line voltage (100 V/120 V, use 5 A/250 V
slow blow; 220 V/240 V, use 2.5 A/250 V slow blow) and reinstall the fuse
compartment by pushing it back into place until the tab locks.
How to Connect to Line Power
Warning
To prevent possible electrical shock, fire, or personal injury:
•
Connect an approved three-conductor mains power cord to
a grounded power outlet.
•
Make sure that the Product is grounded before use.
•
Do not use an extension cord or adapter plug.
The calibrator is shipped with the appropriate line power plug for the country of
purchase. If you need a different type, refer to Table 2-2 and Figure 2-2 for a list and
illustration of the line power plug types available from Fluke.
After you verify that the line voltage selection is set correctly and that the correct fuse for
that line voltage is installed, connect the calibrator to a properly grounded three-prong
outlet.
How to Select Line Frequency
The calibrator is shipped from the factory for nominal operation at 60 Hz line frequency.
If you are using 50 Hz line voltage, you should re-configure the Calibrator for optimal
performance at 50 Hz. To do so, from the front panel, go into SETUP, INSTMT SETUP,
OTHER SETUP, and then push the softkey under MAINS to change the selection to
50 Hz. Store the change. After the instrument is properly warmed up (on for 30 minutes
or longer), you must re-zero the complete instrument. For details, see the section on
“Zeroing the Calibrator” in Chapter 4.
Summary of Contents for 5522A Operators
Page 4: ......
Page 14: ...5522A Operators Manual x ...
Page 18: ...5522A Operators Manual xiv ...
Page 22: ...5522A Operators Manual 1 2 ...
Page 46: ...5522A Operators Manual 1 26 ...
Page 48: ...5522A Operators Manual 2 2 ...
Page 54: ...5522A Operators Manual 2 8 ...
Page 56: ...5522A Operators Manual 3 2 ...
Page 190: ...5522A Operators Manual 6 2 ...
Page 240: ...5522A Operators Manual 7 2 ...
Page 260: ...5522A Operators Manual 8 2 ...
Page 314: ...5522A Operators Manual 9 52 ...
Page 370: ...5522A Operators Manual 10 56 ...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 414: ...5522A Operators Manual B 2 ...
Page 416: ...5522A Operators Manual B 4 ...
Page 426: ...5522A Operators Manual D 6 ...