Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-8
1.
Press
F1
three times to select maximum darkness calibration
Contrast
(CL 0100).
2. Press
F3
(
CALIBRATE)
. The display shows a dark test pattern, see
Figure 5-2.
3. Use
to adjust the display to the maximum darkness where the
test pattern is only just visible.
4.
Press
F3
to return to the softkey menu.
5.
Press
F2
to select default contrast calibration
Contrast (CL 0110):
6. Press
F3
(
CALIBRATE)
. The test pattern shows on the display at default
contrast.
7. Use
to set the display to optimal (this setting becomes the
default) contrast.
8.
Press
F3
to return to the softkey menu.
9.
Press
F2
to select maximum brightness calibration
Contrast (CL 0120):
10. Press
F3
(
CALIBRATE)
. The display shows a bright test pattern.
11. Use
to adjust the display to the maximum brightness where the
test pattern is only just visible.
12. Press
F3
to return to the softkey menu.
13. Now you can either
•
Exit, if only the Contrast had to be adjusted. Continue at the
Save
Calibration Data and Exit
section.
or
•
Do the complete calibration. Press
F2
to select the next step
(Warming-Up) and continue.
hpp204.eps
Figure 5-2. Display Test Pattern