Performance Verification
Scope Input A, B, C, D Tests
4
4-23
Input A AC Voltage Accuracy (HF) and Bandwidth Test
Proceed as follows to test the Input A high frequency automatic scope ac voltage
measurement accuracy and the bandwidth:
1. Connect the Test Tool to the 5502A as shown in Figure 4-6.
2. Select the Test Tool setup:
a) Recall the setup (see
Standard Test Tool Setup
). Press
SAVE
,
F2
(
RECALL) and
select
SETUP
, press
ENTER
, select the setup name, and
press
ENTER
to recall the setup.
b) Press
SCOPE
, then press
F2
(-
READING
…) and select
F1
(
READINGS
)
on
A | V ac
.
c) Press
MANUAL
AUTO
to select autoranging (
AUTO
in upper right LCD edge).
d) Use
mV
RANGE
and
V
RANGE
to change the sensitivity range to select manual
sensitivity ranging, and lock the Input A sensitivity range on 500 mV/div.
AUTO
in upper right LCD edge becomes
½ AUTO
.
e) Use
MOVE
to move the Input A trace zero to the center grid line.
3. Set the 5502A to source a sine wave and to the first test point in Table 4-10.
4. Observe the Input A reading and check that it is within the range shown
under the appropriate column.
5. Continue through the test points.
6. When you are finished, set the 5502A to Standby.
Table 4-10. HF AC Voltage Verification Points
UUT
Model
5502A SC... MODE levsine
UUT
Reading A
Voltage Frequency
All
2.545 Vpp
1 MHz
835 mV to 965 mV
All
2.545 Vpp
25 MHz
790 mV to 1.010 V
190-062
2.545 Vpp
60 MHz
>630 mV
190-104, -102
2.545 Vpp
100 MHz
>630 mV
190-204, -202
2.545 Vpp
200 MHz
>630 mV
190-502; -504
2.545 Vpp
500 MHz
>630 mV