Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-24
Multimeter Meter Zero
Note
The adjustment steps for the meter section are only for the models
190-062, 190-102, and 190-202. For 4-Channel ScopeMeters, go to
the
Save Calibration Data and Exit
section.
Proceed as follows to do the Multimeter (DMM) Zero calibration:
1. Press
F2
to select calibration adjustment step CL 0890.
2. Short circuit (interconnect) the banana jack Meter inputs. Use a test lead
as
short as possible
.
3. Press
F3
to start the zero calibration.
4. Wait until the display shows the status
:READY
.
5. Remove the input terminations.
6. Continue at the
Multimeter (DMM) Volt Gain
section.
Multimeter Volt Gain
Warning
To prevent possible electrical shock, fire, or personal injury,
ensure that the calibrator is in standby mode before making any
connection between the calibrator and the Test Tool.
Dangerous voltages are present on the calibration source and
connection cables during these steps.
Note
The adjustment steps for the meter section are only for the models
190-062, 190-102, 190-202, and 190-502.
To do the DMM Volt Gain calibration:
1. Press
F2
to select the first calibration step in Table 5-7.
2. Connect the Test Tool to the 5502A as shown in Figure 5-9.