Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-6
How to Start the Calibration
To start the calibration:
1. Power the Test Tool with the power adapter input and the BC190 power
adapter.
2. Check the actual Test Tool date and adjust the date if necessary (the
calibration date will become the Test Tool date when saving the calibration
data):
a. Press
USER
(toggles the menu bar on and off).
b. Press
F1
to open the
OPTIONS
menu.
c. Use
to select the
DATE
ADJUST
... option.
d. Press
ENTER
to open the
DATE
ADJUST
menu.
e. If necessary, adjust the date with
. Press
ENTER
to activate all
selections and leave the menu.
3. Select the calibration mode.
The Calibration Adjustment Procedure uses built-in calibration setups that
can be accessed in the calibration mode.
To enter the calibration mode proceed as follows:
•
Press and hold
USER
, press and release
CLEAR
, and release
USER
.
The display shows the
CAL MODE
(Calibration Adjustment) screen.
The display shows the calibration step
WarmingUp (CL 0200)
, the calibration
status
:IDLE (valid)
or
:IDLE (invalid)
, and the softkey menu.
Continue as indicated in the
Calibration Procedure Steps
section.
You can exit the calibration mode without changing the calibration data by turning
the Test Tool off.