Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
2-10
Meter Input (Banana Jacks)
Input Coupling ......................................................... DC
Frequency Response .............................................. DC to 10 kHz (-3 dB)
Input Impedance ..................................................... 1 M
Ω
(
±
1 %)//14 pF (
±
1.5 pF)
Max. Input Voltage ........................................... 1000 V CAT III
600 V CAT IV
(For detailed specifications, see “Safety”)
Meter Functions
Ranging ................................................................... Auto, Manual
Modes ..................................................................... Normal, Relative
General
DC Common Mode Rejection (CMRR) ................... >100 dB
AC Common Mode Rejection at 50, 60,
or 400 Hz ....................................................................... >60 dB
Ohms (
Ω
)
Ranges .................................................................... 500.0
Ω
, 5.000 k
Ω
, 50.00 k
Ω
, 500.0 k
Ω
, 5.000 M
Ω
, 30.00 M
Ω
Full Scale Reading
500
Ω
to 5 M
Ω
.................................................... 5000 counts
30 M
Ω
................................................................. 3000 counts
Accuracy .................................................................
±
(0.6 % +6 counts)
Measurement Current ............................................. 0.5 mA to 50 nA,
±
20 % decreases with increasing ranges
Open Circuit Voltage ............................................... <4 V
Continuity (CONT)
Beep ........................................................................ <50
Ω
(
±
30
Ω
)
Measurement Current ............................................. 0.5 mA,
±
20 %
Detection of shorts of ..............................................
≥
1 ms
Diode
Maximum Voltage Reading ..................................... 2.8 V
Open Circuit Voltage ............................................... <4 V
Accuracy .................................................................
±
(2 % +5 counts)
Measurement Current ............................................. 0.5 mA,
±
20 %
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°
C or
°
F) ................................................... -40.0 to +100.0
°
-100.0
to
+250.0
°
-100.0
to
+500.0
°
-100 to +1000
°
-100 to +2500
°
Probe Sensitivity ..................................................... 1 mV/
°
C and 1 mV/
°
F