Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
5-42
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
USE STACKABLE TEST LEADS
AS SUPPLIED WITH CALIBRATOR
A
B
METER
2 CHANN. + METER
Perf-Ver-Z2.eps
Figure 5-18. Four-wire Ohms Calibration Connections
3. Set the 5502A to the first test point in Table 5-20. Use the 5502A “COMP 2
wire” mode for the calibration adjustments up to and including 100 k
Ω
. For
the higher values, the 5502A will turn off the “COMP 2 wire” mode.
4. Set the 5502A to operate (OPR).
5. Press
F3
to start the calibration.
6. Wait until the display shows the calibration status
:READY
.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points.
8. When you are finished, set the 5502A to Standby.
9. Continue at the
Save Calibration Data and Exit
section.
Table 5-20. Ohm Gain Calibration Points
Cal Step
UUT input Value (5500 NORMAL)
CL 0910
100
Ω
CL 0911
1 k
Ω
CL 0912
10 k
Ω
CL 0913
100 k
Ω
CL 0914
1 M
Ω
CL 0915
10 M
Ω