Calibration Adjustment
Final Calibration For v11.10 and later
5
5-23
4. Set the 5502A to operate (OPR).
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY
.
7. Press
F2
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points of Table 5-6.
8. Set the 5502A to Standby, and continue at the
Multimeter (DMM) Meter Zero
section.
Table 5-6. Input ABCD Gain Calibration Points
Cal step
UUT input value (5502A NORMAL, dc output)
CL 0799
5 mV
CL 0800
12.5 mV
CL 0801
25 mV
CL 0802
50 mV
CL 0803
125 mV
CL 0804
250 mV
CL 0805
500 mV
CL 0806
1.25 V
CL 0807
2.5 V
CL 0808
5 V
CL 0809
12.5 V
CL 0810
25 V
CL 0811
50 V (set 5502A to OPR!)
CL 0812
125 V
CL 0813
250 V