9 - 6
91/92/96/99/105
Users Manual
6.
Use
to highlight 10:1, and press
to
select.
7.
Use
to select CALIBRATE PROBE on A, and
press
to
calibrate.
The test tool outputs a voltage on the GENERATOR OUT
banana jack to the connected probe.
8.
Adjust the trimmer screw in the probe housing of
INPUT A to give an optimum square wave within the
equal markers on the display (see Figure 9-3).
Figure 9-3 Adjusting the Probe
9.
Press
to
continue.
The test tool automatically calibrates itself to the probe.
A message appears on the display indicating that
calibration is successfully completed.
10.
Press
to close the procedure.
INPUT B Probe Calibration
1.
Connect the INPUT B BNC input to the GENERATOR
OUT red banana jack. See Figure 9-4. Use the grey
scope probe and the grey 4-mm banana adapter.
Figure 9-4 INPUT B Probe Calibration Setup
2.
Repeat steps 2 trough 10 of the INPUT A calibration.
91