6 - 20
91/92/96/99/105
Users Manual
Figure 6-12 Startup Menu Window
2. Specify on or off for each of the following:
-
whether or not to identify the waveform (trace)
source (TRACE IDENTIFIERS).
-
whether or not to identify the trigger source, slope,
level, and delay (TRIGGER IDENTIFIER)
-
whether or not to CLEAR at RUN. When Clear at
Run is set to on, and you perform a single shot, the
old waveform is cleared first each time the test tool
writes a new waveform.
-
whether or not to select high or low INFORMATION
LEVEL facility. When the Information Level is set to
high, you will get all on-line information available in
your ScopeMeter test tool.
Factory defaults place all four of these items on.
3. Press
to
select
the
new
Default
Settings.
4. Press
to close the window.
ALTERING THE CONTINUOUS AUTO SET
CONFIGURATION
When
is pressed, the instrument assumes the
Continuous Auto Set configuration.
There are ways to alter the Continuous Auto Set configura-
tion. These two methods are discussed below.
Using Manual Override
Pressing
disables autoranging for voltage attenua-
tion only. Press
to resume the normal mode of opera-
tion for Continuous Auto Set.
Pressing
disables autoranging for the time base
setting only. Press
to resume the normal mode of
operation for Continuous Auto Set.
Changing any of the trigger settings disables autoranging for
voltage, time base, and triggering. Press
to resume
the normal mode of operation for Continuous Auto Set.
Changing Continuous Auto Set Options
Change the Continuous Auto set configuration as follows:
1. From the User Options menu, press
(MORE
OPTIONS) and use
to highlight AUTO SET.
Press
to open the window (see Figure 6-13).