Chapter 4: Maintenance
Autosampler Module Maintenance
96
CyTOF XT User Guide
7
Use the up and down arrows until the end of the probe line is approximately 1
–
2 mm
above the bottom of the tube.
8
Click
Save New Lower Position
after the correct probe line position is achieved.
NOTE
Use the Autosampler camera to verify the correct location of the probe line. The
images below are as examples only.